Normally these commands are used only for spectra line observations. The
bias angle in POLSWITCH 3 and 4 would usually be set to
90
(+ for sources
with DEC < latitude), in which case averaging the results of
POLSWITCH 1 and 3
(or 2 and 4) will correct the Stokes parameter Q (or U) for an instrumental
effect. An observation not position switched is
begun by typing START PLSW
after entering one of the above POLSWITCH modes. The data for each
observation are stored as three scans: the first is the calibrated switched
spectrum (e.g. North minus East), the second and third
are uncalibrated
spectra corresponding to the two individual polarization positions (e.g. the
second scan is North and the third East). The switching rate is set by the
keyword CPER
(the polarimeter will switch positions each CPER seconds).