The CAL routine used for data calibration requires a calibration chopper.
Typing START CAL initiates measurement of the
voltages on the ambient load, the sky, and with the receiver blanked, which
are then used to generate a calibration scan by
computing T
in each
individual channel of the filterbanks. The
cal scan is stored in the T4 register of
the disk data storage area and can be examined via the SPA data reduction
program by typing T4 PLOT at the TEK4010 terminal. A total power
measurement of Tsys
is also computed via the DVM and is displayed on screen 1.
All future data scans are calibrated either with the channel-by-channel (T4)
calibration scan by typing the keyword CALI (brate)
or with the single T
value
displayed on screen 1 by typing the keyword RAW.
If, for some reason CAL
fails one can enter T
manually using the keyword TSYS
( Kelvin) and
observe in RAW mode.
For SEQUOIA, broad band system temperatures are calculated for each pixel and displayed on Screen 15 only if the QEF program is running on the QEF PC in the annex. If this is not running, then the one system temperature is calculated from the DVM which is connected to a spare IF cable from pixel 11 (the center pixel). This value is displayed on screen 1 and screen 15 for each pixel. As long as one is using channel by channel calibration (CALI mode), in which case the calibration is performed within each filter bank, then the state of the QEF PC for calibration purposes is irrelevant. For the FAAS system, the spectra are calibrated by the broad band QEF measurements in RAW mode and by the correlator system temperatures displayed on SCREEN 15 WINDOW 3 in CALI mode.
At some observing frequencies, most notably at
CO (1-0), a significant
correction factor to the computed Tsys in CAL may be required
(see FCRAO
report
220) to accurately calibrate your data.
The magnitude of this factor
depends on the difference between the chopper temperature and effective
atmospheric temperature, T, and the effective atmospheric opacity.
(For example:
CO (1-0) observations through
3 airmasses requires a correction factor of
1.3 for zenith = 0.4 and T = 20 K.) If you can make a reasonable guess at
the zenith opacity and T, enter them via
the keywords TAUZ and DELT and
use the keyword CFAC (ON)
to turn on computation of the correction factor
which will be updated on screen 1 at the beginning of each scan, and will
multiply the filterbank data prior to storage. If you don't use the
correction option [i.e. CFAC (OFF)] then a standard
source should be monitored vs
airmass to determine your own correction curve vs El.
If you wish to correct your data to a standard scale of T
(T
/
) it is
suggested that you use
=
as given in FCRAO report
220 or its successor.
truein